Welcome to the new version of CaltechAUTHORS. Login is currently restricted to library staff. If you notice any issues, please email coda@library.caltech.edu
Published March 1956 | Published
Journal Article Open

Modification to the Automatic Ionization Chamber

Abstract

The automatic ionization chamber has been described previously [1,2]. It has proven quite reliable in service and has many desirable features, especially for balloon-borne equipment. A disadvantage has been that the time of discharge was dependent upon the potential applied to the quartz fiber. Thus, during the comparison of the instrument to be used with the standards, careful measurement of the potential used on the standards was required. Furthermore, care was required to make sure that the battery used during the flight remained constant in potential. It has now been found that the time between rechargings can be made nearly independent of the potential of the fiber by inserting a metal conductor inside and insulating it from the outer conductor surrounding the quartz fiber, and making the potential of this inner conductor the same as that of the fiber.

Additional Information

Copyright © 1956 American Institute of Physics. Received December 27, 1955. Assisted by the joint program of the Office of Naval Research and U.S. Atomic Energy Commission. Reproduction in whole or in part is permitted for any purpose of the United States Government.