Deep reflection-mode photoacoustic imaging and resolution scalability with depth
- Creators
- Song, Kwang Hyun
-
Wang, Lihong V.
Abstract
A deep reflection-mode photoacoustic (PA) imaging system was designed and implemented to visualize deep structures in biological tissues. To achieve good penetration depth, we chose near IR laser pulses at 804 nm wavelength for the generation of photoacoustic waves. To avoid overshadowing the deep PA signals by the surface PA signals, we employed dark-field illumination. To achieve good lateral resolution, we chose spherically focused high-numerical-aperture ultrasonic transducers with 5 MHz or 10 MHz center frequencies. By using these transducers, we achieved 153 μm and 130 μm axial resolutions, respectively, at 19.5 mm depth in 10% porcine gelatin containing 1% intralipid. The system was applied to imaging internal organs of small animals. Compared with our previous high-frequency (50-MHz) photoacoustic microscope, we scaled up the imaging depth while maintaining the ratio of the imaging depth to axial resolution more than 100. In addition, we studied the scalability of the imaging depth and the resolution with ultrasound frequency.
Additional Information
© 2007 Society of Photo-Optical Instrumentation Engineers (SPIE). We thank Konstantine Maslov, Xinmai Yang for fruitful laboratory assistance. The project sponsored by National Institutes of Health grants R01 CA106728 and R01 NS46214 (BRP).Attached Files
Published - 643722.pdf
Files
Name | Size | Download all |
---|---|---|
md5:e3a9a5be2b52d4cf2e8aaa3ddee67b2f
|
604.0 kB | Preview Download |
Additional details
- Eprint ID
- 89827
- Resolver ID
- CaltechAUTHORS:20180921-100901651
- NIH
- R01 CA106728
- NIH
- R01 NS46214
- Created
-
2018-09-27Created from EPrint's datestamp field
- Updated
-
2021-11-16Created from EPrint's last_modified field
- Series Name
- Proceedings of SPIE
- Series Volume or Issue Number
- 6437