Published April 1964
| public
Journal Article
Strontium Contamination in Mineral Analyses
- Creators
-
Wasserburg, G. J.
- Wen, T.
- Aronson, J.
Chicago
Abstract
Significant amounts of strontium contamination occur in analyses of minerals by standard procedures using laboratory glassware. The contamination is much higher than that obtained in blank runs and is associated with the presence of some mineral samples. The contamination can be eliminated by using only teflon and platinum ware during chemical treatment.
Additional Information
© 1964 Published by Elsevier Ltd. The authors wish to acknowledge valuable discussions with Dr. D. Maynes and Dr. Paul Gast.Additional details