Welcome to the new version of CaltechAUTHORS. Login is currently restricted to library staff. If you notice any issues, please email coda@library.caltech.edu
Published March 4, 2013 | Published
Book Section - Chapter Open

Parallel acoustic delay lines for photoacoustic tomography

Abstract

Achieving real-time photoacoustic (PA) tomography typically requires massive ultrasound transducer arrays and data acquisition (DAQ) electronics to receive PA waves simultaneously. In this paper, we report the first demonstration of a photoacoustic tomography (PAT) system using optical fiber-based parallel acoustic delay lines (PADLs). By employing PADLs to introduce specific time delays, the PA signals (on the order of a few micro seconds) can be forced to arrive at the ultrasonic transducers at different times. As a result, time-delayed PA signals in multiple channels can be ultimately received and processed in a serial manner with a single-element transducer, followed by single‐channel DAQ electronics. Our results show that an optically absorbing target in an optically scattering medium can be photoacoustically imaged using the newly developed PADL-based PAT system. Potentially, this approach could be adopted to significantly reduce the complexity and cost of ultrasonic array receiver systems.

Additional Information

© 2013 Society of Photo-Optical Instrumentation Engineers (SPIE).

Attached Files

Published - 85811C.pdf

Files

85811C.pdf
Files (837.8 kB)
Name Size Download all
md5:94a07af90e16d5be06ec8999650a91f8
837.8 kB Preview Download

Additional details

Created:
August 19, 2023
Modified: