Integration and evaluation of a near-infrared camera utilizing a HgCdTe NICMOS3 array for the Mt. Palomar 200-inch Observatory
Abstract
In this paper we describe the main subsystems that constitute the Mt. Palomar Prime Focus InfraRed Camera (PFIRCAM), together with some of the characterization data obtained for the focal plane array. This camera is currently a facility instrument at the 200-inch Mt. Palomar Observatory. It helps to satisfy the observational needs of astronomers in the spectral range of 1 micrometers to 2.5 micrometers by utilizing a HgCdTe NICMOS3 array. The camera has a plate scale of 0.54 arcsec/pixel for an overall FOV of 138 X 138 arcsec.
Additional Information
© 1993 Society of Photo-Optical Instrumentation Engineers (SPIE). The research described in this paper was carried out by the Jet Propulsion Laboratory, California Institute of Technology, under a contract with the National Aeronautics and Space Administration. Reference herein to any specific commercial product, process, or service by trade name, trademark, manufacturer, or otherwise, does not constitute or imply its endorsement by the United States Government or the Jet Propulsion Laboratory, California Institute of Technology.Attached Files
Published - 534.pdf
Files
534.pdf
Files
(1.2 MB)
Name | Size | Download all |
---|---|---|
md5:3119322e889306ef7494b64b340885e4
|
1.2 MB | Preview Download |
Additional details
- Eprint ID
- 88462
- Resolver ID
- CaltechAUTHORS:20180801-133516970
- NASA/JPL/Caltech
- Created
-
2018-08-01Created from EPrint's datestamp field
- Updated
-
2021-11-16Created from EPrint's last_modified field
- Caltech groups
- Infrared Processing and Analysis Center (IPAC)
- Series Name
- Proceedings of SPIE
- Series Volume or Issue Number
- 1946